TherOzone T-8000 Directions Video Demonstration of T-8000 Unit SUMMARY OF OPERATIONS TROUBLESHOOTING • CLEANING THE T-8000 • LIMITED WARRANTY CHEMetrics K-7402 - Aqueous Ozone Testing Kit Directions T-8000 Coupler Lubrication Instructions - (pdf version) • T-8000 "O" Ring Replacement Instructions TherOzone Safety Information • T-8000's Time Clock 2-Page Printable Version of T-8000 Directions (pdf) Notice: Failure to follow these recommended steps or directions for your TherOzone machine may void your warranty. 1. It is extremely important that you do not overfill the water bottles with water *, see fill line instructions below. 2. Never transport the T-8000 unit with the water bottle attached. 3. Do not store the T-8000 unit on its side with the wate bottle attached, especially if the bottle has water in it. 4. Please note: The colder the water the longer the aqueous ozone half life. |
|||
|
* WATER BOTTLE "MAXIMUM FILL LINE"
![]() |
* WATER BOTTLE "MAXIMUM FILL LINE"
(close-up)
![]() |
||
|
STEP ONE
![]() |
STEP
TWO: Filling Bottle
![]() |
![]() |
|
|
Caution:
Before plugging unit into
wall outlet, first place plug into back of T-8000 TherOzone unit. Then
place the plug into the outlet and turn switch on (on back of unit).
|
Fill TherOzone Pressure
Bottle with DISTILLED WATER ONLY
(for highest aqueous ozone concentration). Use 2-4° C or
36-38° F water for both optimum concentration and longest half life
of the aqueous ozone.
|
||
|
STEP
THREE: Maximum Fill Line
![]() |
STEP
FOUR: Connecting TherOzone Pressure Bottle
![]()
|
||
|
CAUTION: Do not exceed maximum
fill line on TherOzone Pressure Bottle. This will void
warranty if overfilled.
|
Lift bottle up towards the
securing male coupler and line up red dots. Stabilize top of
T-8000 with one hand. Turn the bottle clockwise 1/4 turn.
The red dot on the bottle will disappear.
|
||
|
STEP
FIVE: Start Ozonation
![]() |
STEP
SIX: Process is Complete
![]() |
||
|
QUICKLY press the START
button to begin process. Do not hold the START
button down.
|
At approximately 5 minutes,
15 seconds the blue light will light up and the ozonation of the water
will be complete.
|
||
|
STEP
SEVEN: Removing Bottle
![]() |
STEP
EIGHT: Removing Bottle
![]() ![]() |
||
|
To remove bottle turn bottle
counterclockwise and line up the red dots. While supporting
bottom of the bottle with one hand squeeze fingers between bottle
coupler and the T-8000 male coupler.
|
As coupler releases pull
bottle down and away. Note: This is
the only time a slight smell of ozone will be detected during the the
T-8000 operation cycle.
|
||
|
STEP
NINE: Using Ozonated Water
![]() |
GREEN: Power light, unit is plugged in and on/off switch is on. YELLOW: Bottle is in place, coupler is correct, and start button is pressed. BLUE: Cycle is complete and water is ozonated. RED: Light is Off - bottle is in place Light is On - bottle is
not in
place.
Flashing
Red Light
-
pressure in system has dropped. (See
summary of operations)
|
![]() |
|
|
Pour ozonated water from the
TherOzone Pressure Bottle into a dental unit waterline bottle for your
specific dental unit. CAUTION: Do not remove or loosen female coupler from
the TherOzone Pressure Bottle.
|
|||
|
|
|||
SUMMARY OF OPERATIONS • TherOzone
T-8000 unit is plugged in, On switch is pressed, green and red light
are on. NOTE:
If bottle is in place prior
to turning unit on the T-8000 will
not start.
• The TherOzone Pressure Bottle is filled below the Maximum Fill Line with distilled water. This line is below the neck of the bottle. If the bottle is overfilled it will void your warranty. • Bottle is connected, green light is on, and red light is off. Quickly press the start button, yellow and green lights are now on. • Cycle will complete in 5 minutes, 15 seconds, green and blue lights are on. • You may restart cycle with bottle still in place by pressing reset button, then pushing the start button, the cycle will re-run another 5 minutes, 15 seconds. • You may stop cycle by disconnecting the bottle (see Step #9) or by pressing the reset button. • Flashing Red LED - T-8000 has stopped which indicates a pressure leak. Check bottle connection or the interface between the base unit and the coupler. Remove bottle (steps 7 & 8) and reconnect to the unit (step 4). • Flashing Blue and Red LED - T-8000 has stopped - START button was depressed too long. Press RESET to clear. • Blue LED On - If Blue LED comes on soon after pressing the START button, the START button was depressed too long. Press reset to clear. • Your T-8000 will come already lubricated and you will not necessarily need to apply lubrication during its initial phase of use. However if at any
time the,
(1) TherOzone Pressure Bottle does not easily slide on or off the male coupler, or (2) if the turning of the bottle to seal the coupler does not easily glide in either direction, then the silicone lubrication will need to be applied as described on this page, click here "Instructions for Lubricating the T-8000 Coupler." (3) If you have lubricated the T-8000 and it still does not easily glide on and off or seal and be removed easily it is likely that the "O" rings on the coupler will need to be replaced. Click here for Instructions for Replacing the "O" Rings for the T-8000 CLEANING THE T-8000
• Internal components of the unit are
self cleansing due to the inherent properties of the T-8000.
External surfaces should be wiped down with cotton pads saturated with a fresh batch of aqueous ozone (use 2X2" cotton pad bleached with hydrogen peroxide for best results). Do not use harsh chemicals (e.g., bleach, ammonia, etc.). • Do not clear off silicone lubricant from coupler or orange "O' rings. • Your T-8000 will come already lubricated and you will not necessarily need to apply lubrication during its initial phase of use. However if at any
time the,
(1) TherOzone Pressure Bottle does not easily slide on or off the male coupler, or (2) if the turning of the bottle to seal the coupler does not easily glide in either direction, then the silicone lubrication will need to be applied as described on this page, click here "Instructions for Lubricating the T-8000 Coupler." • This warranty does not apply to any products that have been mishandled, misused, neglected or modified in any way, according to the TherOzone, LLC directives. This includes tamper resistant seals that show any evidence of manipulation, any "O' rings that have been stripped of protective silicone lubricant and any improperly cleansed surfaces. • TherOzone, LLC retains the right to repair or rebuild returned products. All refunds are solely at the discretion of TherOzone, LLC. • TherOzone, LLC retains sole authorization of warranty to original purchaser of record. Transfer of warranty is at the discretion of TherOzone, LLC. • The limited warranty is good for 1-year from the date of purchase or 300 hours of operation, determined by the clock on the back of the T-8000 unit (whichever comes first). • For further questions please see contact information below and be sure to follow step by step Directions for Use of T-8000 Unit. • To test if your T-8000 unit is optimally functioning and producing aqueous ozone 3.2-4.0 ppm you will need to purchase a CHEMetrics K-7402 Tester and follow TherOzone's CHEMetrics Directions for properly diluting and calculating aqueous ozone content. |
|||